Z-1633-2026

Class II · Ongoing

Recalling firm
Straumann USA LLC
Classification
Class II
Status
Ongoing
Initiated
20260317
Firm location
Andover, MA, United States

Product description

Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Reason for recall

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Data sourced from openFDA. This site is unofficial and independent of the FDA.